3001 Levelogger 5 LTC, M5/C80
3001 Levelogger 5 LTC, M5/C80
3001 Levelogger 5 LTC, M5/C80
3001 Levelogger 5 LTC, M5/C80
Solinst

3001 Levelogger 5 LTC, M5/C80

Pine Item #52481
The Levelogger 5 LTC includes memory for 100,000 readings of level, temperature, and conductivity, and a 8-year lithium battery. The 22 mm x 208 mm (7/8" x 8.2") stainless steel body has a baked-on finish (inside and out) using polymerization technology for extra corrosion, abrasion, and chemical resistance.
Levelogger Software includes a Calibration Wizard that guides the user through a 1 to 4-point calibration of the conductivity sensor, using standard solutions. The conductivity sensor autoranges from 0–100,000 µS/cm and calibrates from 50–80,000 µS/cm. Accuracy is ±1% from 5,000 µS/cm–80,000 µS/cm and the greater of ±2% or 15 µS/cm from 50 µS/cm–5,000 µS/cm.
The Levelogger 5 LTC can be set to record every 2 seconds to 99 hours, in linear, event-base, or scheduled sampling modes. It is compatible with all Levelogger communication and deployment accessories. A Barologger 5 can be used to compensate the water level readings.
Features:

Large memory: 100,000 sets of data

Stable communication: single-eye optical, easy to clean, more scratch resistant

Strong, robust design: double o-ring seals for two times over pressurization rating

Enhanced thermistor and conductivity sensitivity: accurate platinum RTD and conductivity sensor

Superior protection in harsh conditions: baked-on coating using polymerization—inside and out

Applications:

Salt water intrusion and soil salination monitoring

Plume remediation monitoring and studies

Leachate monitoring at landfills, mine tailings, waste disposal storage sites, and more

Agricultural and stormwater runoff monitoring

Create a historical database for potable water supply monitoring

Tracer tests

Quantity must be 1 or more
Equipment available for both rental and purchase
Request a Quote

Specifications

Specifications

Level Sensor:
Ranges:
Accuracy:
Resolution:
Units of Measure:
Normalization:
Temp. Comp. Range:
Piezoresistive Silicon with Hastelloy Sensor
5, 10, 20, 30, 100, and 200 m
±0.05% FS
0.001% FS to 0.0006% FS
cm, m, ft, psi, kPa, bar (ºC, ºF)
Automatic Temperature Compensation
0º to 50ºC
Temperature Sensor:
Accuracy:
Resolution:
Platinum Resistance Temperature Detector (RTD)
±0.05ºC
0.003ºC
Conductivity Sensor:
Full Range:
Calibrated Range:
Accuracy:
Resolution:
Temp Comp. Range:
Normalization:
4-Electrode Platinum
0 – 100,000 µS/cm
50 – 80,000 µS/cm
±1%: 5,000 µS/cm – 80,000 µS/cm;
greater of ±2% or 15 µS/cm: 50 µS/cm –5,000 µS/cm
±0.1 µS/cm
0ºC – 50ºC
Specific Conductance @ 25˚C
Battery Life:
Clock Accuracy (typical):
Operating Temperature:
8 Years (1 reading every 5 minutes)
±1 minute/year (-20ºC to 80ºC)
-20ºC to 80ºC
Maximum Readings:
Memory:
Communication:
100,000 sets of readings
Slate or Continuous
Optical high-speed: 57,600 bps with USB
Size:
Weight:
Corrosion Resistance
Wetted Materials:
22 mm x 208 mm (7/8" x 8.2")
200 grams (7.06 oz)
Baked-on coating using polymerization
Platinum, Delrin®, Viton®, 316L Stainless Steel,
Hastelloy, Regulator approved PFAS-free PTFE
(inside and out)
Sampling Mode:
Measurement Rates:
Barometric Compensation:
Linear, Event & User-Selectable with Repeat Mode,
Future Start, Future Stop, Real-Time View
2 seconds to 99 hours
Software Wizard and Barologger 5